SFE-GC with quantitative transfer of the extraction effluent to a megabore capillary column
1Virginia Polytechnic Institute and State University, Department of Chemistry, Blacksburg, 24061-0212, USA.
Abstract:
Supercritical fluid extraction was directly coupled with gas chromatography such that the entire extraction effluent was directed onto the GC column. A sealed connection between the restrictor and the column ensured that no loss of analyte could occur. The elevated pressure generated in the column itself resulted in a reduced flow rate through the column, making efficient trapping possible. Metal-jacketed capillary columns were used as they can tolerate the elevated pressures required. In addition to maximizing sensitivity, very practical extraction parameters were possible. For a 25-min extraction, with a 2.2 mL/min flow rate, and the column at 40 degrees C, analytes as volatile as C-14 were trapped and separated. Trapping efficiency improved further with shorter extraction times, mild cooling of the column, or lower flow rates.
More Related Videos
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
High-Performance Liquid Chromatography: Elution Process
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...