Intensity modulated radiotherapy dose delivery error from radiation field offset inaccuracy.
Medical Physics
|August 18, 2000
Summary
Accurate multileaf collimator calibration is crucial in Intensity Modulated Radiotherapy (IMRT). A Radiation Field Offset (RFO) error of 0.5 mm can cause significant dose errors in target volumes, necessitating precise calibration for effective cancer treatment.
Area of Science:
- Medical Physics
- Radiation Oncology
Background:
- Intensity Modulated Radiotherapy (IMRT) relies on precise multileaf collimator (MLC) calibration.
- Inaccuracies in MLC calibration, specifically the Radiation Field Offset (RFO), can lead to significant dose errors within target volumes.
- The CORVUS IMRT system's sensitivity to RFO errors requires thorough investigation.
Purpose of the Study:
- To analyze the dosimetric consequences of incorrect RFO values in the CORVUS step-and-shoot IMRT system.
- To quantify the Dose Error Sensitivity Factor (DESF) for various IMRT plans.
- To establish an analytic framework for understanding and predicting RFO-related dose errors.
Main Methods:
- IMRT plans were generated with correct RFO values for a phantom target volume.
- A computer code was developed to simulate delivery with incorrect RFO values by modifying dynamic MLC (DMLC) files.
- Doses were measured using an ionization chamber after irradiating with both correct and incorrect RFO DMLC files.
Main Results:
- Measured DESF values ranged from 0-8% mm⁻¹, indicating varying sensitivity to RFO errors across different plans.
- For the CORVUS system (6 MV, 1 cm x 1 cm, step-and-shoot IMRT), an RFO accuracy of 0.5 mm is required to maintain a fractional dose error below 4%.
- A strong correlation was found between high DESF values and DMLC files with small average leaf openings, with a predicted maximum DESF of 20% mm⁻¹.
Conclusions:
- Precise RFO determination is critical for accurate dose delivery in step-and-shoot IMRT.
- The developed analytic framework and computer code can predict and screen IMRT plans for RFO sensitivity.
- Ensuring sub-millimeter accuracy in RFO calibration is essential to minimize dose errors in IMRT treatments.