Background:
Preoxygenation before tracheal intubation is intended to increase oxygen reserves and delay the onset of hypoxemia during apnea. Various systems are used for preoxygenation. Designed specifically for preoxygenation, the NasOral system uses a small nasal mask for inspiration and a mouthpiece for exhalation. One-way valves in the nasal mask and the mouthpiece ensure unidirectional flow. This investigation compares the efficacy of preoxygenation using the standard circle system with the NasOral system and five different resuscitation bags.
Methods:
Twenty consenting, healthy volunteers were studied in the supine position for 5-min periods of tidal volume breathing using the circle absorber system, the NasOral system, and five resuscitation bags in a randomized order. Data were collected during room air breathing and at 30-s intervals during 5 min of oxygen administration. Inspired oxygen, end-tidal oxygen, and end-tidal nitrogen were measured by mass spectrometry.
Results:
At 2. 5 min of oxygenation, end-tidal oxygen plateaued at 88.1 +/- 4.8 and 89.3 +/- 6.4% (mean +/- SD) for the circle absorber and NasOral systems, respectively. This was associated with inverse decreases in end-tidal nitrogen. At no time did these end-tidal oxygen or nitrogen values differ from each other. Three of the resuscitation bags (one disk type and two duck-bill type with one-way exhalation valves) delivered inspired oxygen more than 90%, and the end-tidal oxygen plateaued between 77 and 89% at 2 min of tidal volume breathing. The other two resuscitation bags (both duck-bill bags without exhalation valves) delivered inspired oxygen less than 40%, and the end-tidal oxygen values ranged between 21.8 +/- 5.0 and 31.9 +/- 8.7%.
Conclusions:
The circle absorber and NasOral systems were equally effective in achieving maximal preoxygenation during tidal volume breathing. Resuscitation bags differed markedly in effectiveness during preoxygenation; those with duck-bill valves without one-way exhalation valves were the least effective. Thus, the use of these bags should be avoided for preoxygenation.
More Related Videos
08:35Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
Related Concept Videos
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Cardiopulmonary Resuscitation II: ACLS Airway Management