Related Experiment Videos
Respiratory monitoring using an air-mattress system
P Chow1, G Nagendra, J Abisheganaden
1Electronics and Communication Engineering Department, Singapore Polytechnic, Republic of Singapore.
Abstract:
This paper describes a non-invasive respiratory monitoring system using an air mattress. The air-mattress system features multiple air compartments to monitor movements of the thorax and the abdomen separately. To evaluate the performance of the air-mattress system, four subjects were selected for the study and their separate ribcage and abdominal movements were monitored simultaneously by respiratory inductive phlethysmography belts and the air-mattress system. The sensitivity and accuracy indices of the air-mattress system for the detection of hypopnoeas scored remarkably well (above 90%). In addition, it was noted that the mean error in the measurement of the respiratory rate between the two systems was very small. This paper shows that the air-mattress system can be a reliable non-invasive respiratory monitoring system to detect simple abnormalities in breathing, such as respiration rate and hypopnoeas.