Reticulocyte analysis using light microscopy and two different flow cytometric procedures
I Preloznik-Zupan1, P Cernelc, D Zontar
1Department of Haematology, University Medical Centre, Ljubljana, Slovenia.
Pflugers Archiv : European Journal of Physiology
|September 27, 2000
Summary
Flow cytometry offers a more accurate and reproducible method for reticulocyte counts compared to traditional microscopy. This study validates flow cytometry
Area of Science:
- Hematology
- Clinical Pathology
- Biomedical Engineering
Background:
- Reticulocyte counts are crucial indicators of bone marrow erythropoietic activity.
- Microscopic reticulocyte enumeration suffers from inaccuracy and poor reproducibility.
- Flow cytometry offers automated and improved quality for reticulocyte analysis.
Purpose of the Study:
- To determine the optimal thiazole orange (TO) concentration for EPICS Profile flow cytometer.
- To assess the correlation between microscopic and two flow cytometric reticulocyte counting methods.
- To evaluate the clinical utility of flow cytometry for routine reticulocyte analysis.
Main Methods:
- Comparison of three reticulocyte enumeration methods: microscopic (Brilliant Cresyl Blue) and two flow cytometric (FACSort, EPICS Profile) using thiazole orange (TO).
- Optimization of TO concentration for the EPICS Profile system.
- Correlation analysis between the tested methods and assessment of intra-assay variability.
Main Results:
- The optimal TO concentration for the EPICS Profile was found to be 0.1 mg/L.
- Good correlation (coefficients 0.82–0.87) was observed between microscopic and flow cytometric methods.
- Flow cytometry demonstrated significantly lower intra-assay coefficients of variation (6.3%–8.4%) compared to microscopy (27.5%).
Conclusions:
- Flow cytometry, particularly using thiazole orange, provides a reliable and reproducible method for reticulocyte counting.
- The optimized flow cytometric methods are suitable for routine clinical laboratory use.
- Automation via flow cytometry significantly enhances the accuracy and consistency of reticulocyte enumeration.