Phantom study of fusion image of CT and SPECT with body-contour generated from external Compton scatter sources
K Suga1, N Matsunaga, Y Kawakami
1Department of Radiology, Yamaguchi University School of Medicine, Ube, Japan.
Purpose:
A phantom study was conducted to evaluate the feasibility of body contour definition with Compton scatter photons from external sources of technetium-99m pertechnetate (Tc-99m) to create a fusion image of CT and SPECT images.
Methods:
External sources of 1 mCi (37 MBq) Tc-99m were placed on each collimator, and body-contour SPECT images were obtained with an energy window of 100 keV +/- 25% for detecting 90 degrees and 180 degrees Compton scatter photons of Tc-99m from the body surface in water-filled cylindrical and hexagonal phantoms, and in a chest phantom with a Tc-99m-avid simulated lung nodule and multimethod surface markers. In the chest phantom, each transaxial SPECT slices was registered with the corresponding CT slice by using image-matching soft ware. A summation of the registered images yielded a three-dimensional (3-D) fusion image of this phantom.
Results:
This method clearly visualized the body contour on all the SPECT slices in all the phantoms except for the complex hexagonal phantom. There was no significant difference between the known and SPECT-measured diameters of the cylindrical phantom. The fit of CT and SPECT images of the chest phantom was achieved with a mean alignment error of 5% in visual inspection, which was improved to 0.2% after correction of the magnification of the SPECT images according to the resultant dimensional differences. The 3-D fusion image of this phantom effectively visualized the anatomic location of the lung nodule and surface markers.
Conclusion:
This simple method effectively provided boundary information on the cold phantoms. Although further improvements in the registration technique with CT images are desirable, the body-contour SPECT image obtained by this method has the potential for accurately creating a 3-D fusion image with CT images, and is a feasible way of anticipating the anatomical localization of a target tissue.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET