Performance of transport ventilator with patient-triggered ventilation
Chest
|October 18, 2000
Summary
New transport ventilators demonstrate performance comparable to ICU ventilators. These devices, featuring advanced triggering and pressure support, are suitable for clinical use.
Area of Science:
- Mechanical Ventilation
- Respiratory Care Engineering
Background:
- Transport ventilators with advanced features (inspiratory triggering, pressure support-control) are now available.
- Comparison with standard Intensive Care Unit (ICU) ventilators is needed.
Purpose of the Study:
- To evaluate the performance of new transport ventilators against a standard ICU ventilator.
- To assess inspiratory and expiratory functions, and oxygen consumption.
Main Methods:
- Laboratory study using a mechanical lung model simulating spontaneous breathing.
- Compared four transport ventilators (Mallinckrodt 740, Bird TBird, Pulmonetic LTV1000, Respironics Esprit) with a Mallinckrodt 7200ae ICU ventilator.
- Evaluated inspiratory delay time, trigger pressure, pressure rise time, expiratory pressure, time constant, and oxygen requirement under various settings.
Main Results:
- Most transport ventilators exhibited clinically satisfactory inspiratory delay times (<100 ms).
- Expiratory time constants were significantly shorter for transport ventilators compared to the ICU unit.
- Esprit ventilator showed the lowest oxygen requirement, followed by 740, LTV1000, and TBird.
Conclusions:
- Newer FDA-approved transport ventilators show performance comparable to ICU ventilators.
- These transport ventilators are likely suitable for clinical application in critical care settings.