Validation of A&D UA-767 device for the self-measurement of blood pressure
A N Rogoza1, T S Pavlova, M V Sergeeva
1Department of New Diagnostic Methods, Cardiology Research Complex, Moscow, Russia. anrogoza@mtu-net.ru
Background:
The validation of self-measurement devices has been recommended. Automatic monitor A&D UA-767 (A&D Company, Ltd, Tokyo, Japan) is well known and widely used, but not tested according to the Association for Advancement of Medical Instrumentation (AAMI) and British Hypertension Society (BHS) recommendations.
Objective:
To perform a clinical validation for use by adults of the A&D UA-767 device according to the criteria of the AAMI and a modified BHS protocol.
Methods:
The test concerned 101 subjects (57 men and 44 women) aged 15-85 years with arm circumferences of 22-39 cm, a systolic blood pressure (SBP) range of 89-206 mmHg, and a diastolic blood pressure (DBP) range of 53-122 mmHg. For each subject, three readings of the UA-767 were compared with simultaneous auscultatory measurements by two trained independent observers who used a mercury manometer and dual stethoscope. The results were graded according to the BHS (1990 and 1993) and AAMI recommendations.
Results:
Observers showed close agreement, with mean differences of 1.1+/-2.4 mmHg for SBP and -0.7+/-2.0 mmHg for DBP. The proportion of values agreeing to within 5, 10 and 15 mmHg were 93, 100, and 100% for SBP and 97, 100, and 100% for DBP for the two observers. The sphygmomanometer measurements were 132+/-24/79+/-14 mmHg (mean+/-SD). The average difference between the mercury sphygmomanometer and A&D UA-767 readings for SBP and DBP were, respectively, -0.4+/-5.4 and -0.4+/-4. 8 mmHg. The proportion of values agreeing to within 5,10, and 15 mmHg were 82, 94, and 98% for SBP and 80, 95, and 98% for DBP for the observers and device (A/A grade for BHS).
Conclusions:
For an adult population, the A&D UA-767 device for the self-measurement of blood pressure satisfied the AAMI criteria, achieved a BHS grade of A/A and can therefore be recommended for monitoring blood pressure in home and clinical conditions by patients with mild-to-moderate arterial hypertension.
More Related Videos
09:33A Methodological Approach to Non-invasive Assessments of Vascular Function and Morphology
Published on: February 7, 2015
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure