Related Experiment Video
Updated: Jul 18, 2026

Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
Assisted venous drainage presents the risk of undetected air microembolism
A Lapietra1, E A Grossi, B B Pua
1Department of Surgery, Division of Cardiothoracic Surgery, New York University School of Medicine, New York, NY, USA.
Objectives:
The proliferation of minimally invasive cardiac surgery has increased dependence on augmented venous return techniques for cardiopulmonary bypass. Such augmented techniques have the potential to introduce venous air emboli, which can pass to the patient. We examined the potential for the transmission of air emboli with different augmented venous return techniques.
Methods:
In vitro bypass systems with augmented venous drainage were created with either kinetically augmented or vacuum-augmented venous return. Roller or centrifugal pumps were used for arterial perfusion in combination with a hollow fiber oxygenator and a 40-micrometer arterial filter. Air was introduced into the venous line via an open 25-gauge needle. Test conditions involved varying the amount of negative venous pressure, the augmented venous return technique, and the arterial pump type. Measurements were recorded at the following sites: pre-arterial pump, post-arterial pump, post-oxygenator, and patient side.
Results:
Kinetically augmented venous return quickly filled the centrifugal venous pump with macrobubbles requiring continuous manual clearing; a steady state to test for air embolism could not be achieved. Vacuum-augmented venous return handled the air leakage satisfactorily and microbubbles per minute were measured. Higher vacuum pressures resulted in delivery of significantly more microbubbles to the "patient" (P <.001). The use of an arterial centrifugal pump was associated with fewer microbubbles (P =.02).
Conclusions:
Some augmented venous return configurations permit a significant quantity of microbubbles to reach the patient despite filtration. A centrifugal pump has air-handling disadvantages when used for kinetic venous drainage, but when used as an arterial pump in combination with vacuum-assisted venous drainage it aids in clearing air emboli.
Related Concept Videos
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Venous Thrombosis I: Introduction
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Venous Thrombosis III: Interprofessional Care
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care
Pulmonary Embolism III: Nursing Management