Reproducibility--an important factor determining the quality of computer-aided detection (CAD) systems
1Institute of Diagnostic and Interventional Radiology, Friedrich-Schiller-University Jena, Bachstrasse 18, 07740 Jena, Germany. ansgarmalich@gmx.de
European Journal of Radiology
|November 25, 2000
Summary
The reproducibility of computer-aided detection (CAD) markings on mammography films was tested. The current CAD system
Area of Science:
- Radiology
- Medical Imaging
- Artificial Intelligence in Medicine
Background:
- Mammography is a key tool for breast cancer screening.
- Computer-aided detection (CAD) systems aim to improve mammography interpretation.
- Reproducibility of CAD system markings is crucial for clinical adoption.
Purpose of the Study:
- To evaluate the reproducibility of markings generated by a commercial computer-aided detection (CAD) system on mammography films.
- To assess the sensitivity of the CAD system in detecting malignant lesions, including microcalcifications and masses.
Main Methods:
- One hundred unilateral mammography examinations with histopathologically proven malignant lesions were retrospectively scanned three times using a CAD system.
- Reproducibility and sensitivity (strict and broader sense) were determined for microcalcifications and masses.
- The study included cases with suspicious microcalcification clusters, masses, or both.
Main Results:
- Only 18% of mammography examinations showed identical CAD markings across all three scans, indicating low reproducibility.
- The CAD system achieved strict sensitivity of 23.4% for microcalcifications and 51.1% for masses.
- False negative rates were 12.8% for microcalcifications and 9.6% for masses.
Conclusions:
- The reproducibility of the evaluated CAD system is currently insufficient for routine clinical use.
- Improvements in CAD system characteristics are needed to enhance its value as a 'second reader' in mammography interpretation.
- Further development is required to ensure reliable and consistent CAD performance in clinical practice.