Related Experiment Video
Updated: Oct 9, 2026

Determination of Inorganic Arsenic in a Wide Range of Food Matrices using Hydride Generation - Atomic Absorption Spectrometry.
Published on: September 1, 2017
Use of automated solid-phase extraction equipment for the determination of isoniazid in milk
L Grasso1, G Scarano, E Imparato
1Dipartimento di Chimica, Istituto Zooprofilattico Sperimentale del Mezzogiorno (IZSM), Portici, Italy.
Abstract:
The possibility of using automated equipment for solid phase extraction of isoniazid from milk has been investigated. This equipment provides two noteworthy advantages: it shortens the time of analysis and avoids SPE column plugging, stabilizing flow rates. After setting the correct programme on software controlling the equipment, some validation parameters have been determined to obtain recovery and repeatability data. The mean recovery for spiked samples resulted in the range 75-95%, the standard error ranged from -5 to -20% while the RSD varied from 2 to 22%. Therefore the method can be considered reliable and fast.