Sevoflurane output from the Oxford Miniature Vaporizer in drawover mode
1Department of Anaesthesia, National University Hospital, Singapore, Republic of Singapore.
Abstract:
The sevoflurane output from the Oxford Miniature Vaporizer (OMV) was measured in a series of bench tests in a drawover system. Using a range of settings on two OMVs in series, output ranged from 0.5 to 5.9% with a 600 ml tidal volume at a rate of 12 breaths/min produced by a Penlon Multivent. The OMVs' output was steady over 40 minutes. In tests with tidal volumes ranging from 300 ml to 1200 ml, the OMVs' output was lower at higher minute volumes. When continuous flow of carrier gas at 7.2 l/min was used, the output was lower than at the equivalent intermittent flow rate. Using two OMVs with a drawover system appears to be a feasible technique for the induction and maintenance of sevoflurane anaesthesia. Clinical tests of sevoflurane drawover anaesthesia are now required so that the advantages of sevoflurane can gain wider application in field and military anaesthesia.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Vapor Pressure Lowering
Inhalational Anesthetics: Overview
Vaporization