Direct comparison of a xenon and a solid-state CT detector system: measurements under working conditions
T Fuchs1, M Kachelriess, W A Kalender
1Institute of Medical Physics, University of Erlangen-Nürnberg, Erlangen, Germany.
Abstract:
Measurements of various image quality parameters were carried out with two different detector systems in an otherwise unchanged medical computed tomography (CT) scanner. As all other components of the scanner and the image reconstruction system remained identical, we were able to quantify the difference in performance between a Xenon gas ionization detector and a new solid-state scintillation detector in an isolated fashion. We determined noise, spatial resolution, and artifact behavior and assessed the potential for dose reduction. No significant impact of the detector change on absolute CT values of a calibration phantom was observed. Spatial resolution was improved by more than 10% for the solid-state system. As the system's modulation transfer functions were measured with a wire phantom and otherwise unchanged scanner geometry and image reconstruction algorithm, the increase of resolution is explained by the improved temporal response of the solid-state detector. At the same time, noise was reduced by 12% for a 20-cm diameter water phantom. The noise reduction corresponds to a possible reduction of patient dose by 23% for constant image quality, which is in good agreement with our prediction by estimations of both systems total detective quantum efficiency. Also, a significant improvement of scatter rejection was found for the solid-state system.
More Related Videos
08:23Quantitative Measure of Lung Structure and Function Obtained from Hyperpolarized Xenon Spectroscopy
Published on: November 10, 2023
10:42In Depth Analyses of LEDs by a Combination of X-ray Computed Tomography (CT) and Light Microscopy (LM) Correlated with Scanning Electron Microscopy (SEM)
Published on: June 16, 2016
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Atomic Emission Spectroscopy: Instrumentation
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...