Related Experiment Video
Updated: Jul 31, 2026

The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation
Published on: February 28, 2012
Lead- and device-related complications in the antiarrhythmics versus implantable defibrillators trial
1Oregon Health Sciences University, Portland, Ore, Sentara Norfolk General Hospital, Norfolk, VA, USA. AVID Investigators.
Background:
Implantation of transvenous implantable cardioverter defibrillators (ICDs) by use of a nonthoracotomy approach has become routine therapy for survivors of life-threatening tachyarrhythmias. The purpose of this study was to identify and prospectively characterize the frequency of lead- and ICD-related complications from the Antiarrhythmics versus Implantable Defibrillators (AVID) Trial.
Methods And Results:
Between June 1, 1993, and April 7, 1997, 539 patients received nonthoracotomy ICDs either as initial treatment assignment (477) or as crossover from medical management (62). A total of 62 first complications occurred. The subclavian route of insertion resulted in more complications than the cephalic vein route, 46 of 339 (14%) versus 6 of 135 (4%), P = .005, as did the abdominal versus pectoral generator site, 31 of 238 (13%) versus 17 of 291 (6%), P<.02. Most dislodgements and system infections tended to occur in the 3 months after implantation, whereas lead fractures continued to occur throughout follow-up. Failure to use perioperative antibiotics was a predictor of system infection (P = .001).
Conclusions:
These data suggest that cephalic vein access and pectoral generator site may result in fewer complications. The continued occurrence of lead fractures and the need for premature system revision supports the practice of close routine ICD system surveillance.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Mechanism of Cardiac Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias V: Evaluating Dysrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Cardiomyopathy V: Interprofessional Care
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias