Background:
During low-flow or closed-circuit anesthesia with the fluorinated inhalation anesthetic sevoflurane, compound A, an olefinic degradation product with known nephrotoxicity in rats, is generated on contact with alkaline CO(2) adsorbents. To evaluate compound A formation and thus potential sevoflurane toxicity, a reliable and reproducible assay for quantitative vapor-phase compound A determination was developed.
Methods:
Compound A concentrations were measured by fully automated capillary gas chromatography-mass spectrometry with cryofocusing. Calibrators of compound A in the vapor phase were prepared from liquid volumetric dilutions of stock solutions of compound A and sevoflurane in ethyl acetate. 1,1,1-Trifluoro-2-iodoethane was chosen as an internal standard. The resulting quantitative method was fully validated.
Results:
A linear response over a clinically useful concentration interval (0.3-75 microL/L) was obtained. Specificity, sensitivity, and accuracy conformed with current analytical requirements. The CVs were 4.1-10%, the limit of detection was 0.1 microL/L, and the limit of quantification was 0.3 microL/L. Analytical recoveries were 100.6% +/- 10.1%, 102.5% +/- 7.3%, and 99.0% +/- 4.1% at 0.5, 10, and 75 microL/L, respectively. The method described was used to determine compound A concentrations during simulated closed-circuit conditions. Some of the resulting data are included, illustrating the practical applicability of the proposed analytical approach.
Conclusions:
A simple, fully automated, and reliable quantitative analytical method for determination of compound A in air was developed. A solution was established for sampling, calibration, and chromatographic separation of volatiles in an area complicated by limited availability of sample volume and low concentrations of the analyte.
More Related Videos
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Volatilization
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Stages of General Anesthesia
Inhalational Anesthetics: Overview