B A de Mol1, M E Cromheecke, J G Groen
1Department of Cardiothoracic Surgery, Academic Medical Center, University of Amsterdam, Amsterdam, The Netherlands. b.a.demol@amc.uva.nl
Abstract:
Fractures in Björk-Shiley convexoconcave (BScc) heart valves have raised questions about the feasibility of early diagnosis of technical defects by means of acoustic assessment. Three laboratory tests were conducted. To establish acoustic fingerprints, 66 valves with a defect, such as single-leg fracture (SLF) or single-leg separation (SLS), or without a defect were connected with a contact sensor and excited by dropping a small metal ball onto the outlet strut. In the second test, we simulated the valve sound propagation within the thorax. In the third test, intact, SLF, and SLS valves were placed in a mock heart immersed in a large water tank. We observed a resonance frequency corresponding with valve size and presence of defects. The second test showed that both the chest wall and the lungs created numerous reflections. This led to a substantial overlap of the original pulse frequencies and the frequencies measured. The third test confirmed that submersion of the chest in water can significantly reduce chest wall reflections. Reliable noninvasive assessment of BScc valve clicks for the presence of defects of the outlet strut is hampered by complex sound propagation within the thorax and variability of valve excitation. Acoustic fingerprints to diagnose mechanical defects should be integrated in valve design.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.
Cardiovascular System Abnormal Findings II: Auscultation
Abnormal Heart Sounds
Gallops:
Mitral Valve Prolapse I: Introduction
Mitral Regurgitation I: Introduction
Aortic Regurgitation I: Introduction