Related Experiment Videos
Simultaneous measurement of 89Sr and 90Sr in aqueous samples by liquid scintillation counting using the spectrum
1Nuclear Environment Research Division, Korea Atomic Energy Research Institute, Yusong, Taejon, South Korea. khhong@nanum.kaeri.re.kr
Abstract:
Two radiostrontium isotopes, 89Sr and 90Sr, were measured simultaneously by liquid scintillation counter coupling with a full spectrum DPM counting protocol. Strontium was separated from the aqueous media by extraction chromatography using Sr-spec resin. Individual count rates were evaluated from quench correlation curves for 89Sr and 90Sr related with SIS and tSIE quench indicators. The method exhibited a counting efficiency of 95% for 89Sr and 92% for 90Sr within their full energy channels and gave highly precise results when the sample was counted within 4 h after strontium separation. Lower limits of detection were 37 mBq/l for 90Sr and 32 mBq/l for 89Sr with a 60 min counting time.