Fourier transform infrared determination of caffeine in roasted coffee samples
J M Garrigues1, Z Bouhsain, S Garrigues
1Department of Analytical Chemistry, University of Valencia, Spain.
Abstract:
A new procedure has been developed for the FT-IR determination of caffeine in roasted coffee samples. The method involves wetting the coffee samples with a 0.25 M aqueous NH3 solution, extracting the caffeine with CHCl3, and measuring absorbance at 1,659 cm(-1) using a baseline established between 1,900 and 830 cm(-1). The procedure proposed is fast, only requiring a total extraction time of 16 min for each sample, and provides a drastic reduction of the organic solvent consumed, from the 200 mL diethyl ether and 50 mL CHCl3, required for each sample by the reference chromatographic UV-spectrometric determination to only 5 mL CHCl3. The method provides a limit of detection of the order of 3 mg L(-1) caffeine and a relative standard deviation of 0.4% for 3 independent analyses of a sample containing 18.6%mg/g caffeine. The accuracy of the FT-IR procedure was evaluated from recovery experiments on spiked samples providing values from 94.4 to 100.1% and from the comparison of results found for a series of commercial samples, by both FT-IR and the official reference procedure.
More Related Videos
Related Concept Videos
Infrared (IR) Spectroscopy: Overview
Different compounds display unique properties due to their...
IR Spectrum
Transmittance is defined as the ratio of the radiant power passing through a sample to that from the radiation's source. Multiplying the transmittance by 100 gives the percent transmittance (%T), which varies between 100% (no absorption) and 0% (complete...
IR and UV–Vis Spectroscopy of Aldehydes and Ketones
IR Spectrometers
IR Frequency Region: X–H Stretching
IR Frequency Region: Fingerprint Region
The...