Related Experiment Video
Updated: Aug 8, 2026

Phase-Resolved Functional Lung MRI for Pulmonary Ventilation and Perfusion (V/Q) Assessment
Published on: August 9, 2024
Variability in the cardiac EIT image as a function of electrode position, lung volume and body position
R P Patterson1, J Zhang, L I Mason
1Department of Biomedical Engineering, University of Minnesota, Minneapolis 55455, USA. patte001@tc.umn.edu
Abstract:
A study was conducted using the Sheffield electrical impedance tomography (EIT) portable system DAS-01 P to determine the change in the cardiac image with electrode position, lung volume and body position. Sixteen electrodes were positioned in three transverse planes around the thorax at the level of the second intercostal space, at the level of the xiphisternal joint, and midway between upper and lower locations. Data were collected at each electrode level with the breath held at end expiration and after inspiring 0.5, 1 and 1.5 l of air with the subject in both the supine and sitting position. These data were analysed using a Matlab developed program that calculates the average resistivity change in the cardiac region from automatically determined borders. Results show significant individual variability with electrode position and air volume. The middle electrode most consistently shows an increase in impedance in the region of the heart during systole. In some subjects the change in the ventricular-volume-like curve showed a greater than 50% change as a function of lung volume. The pattern of variability with electrode position was not consistent among subjects. In one subject MRI images were obtained to compare actual structures with those seen in the EIT image. The results suggest that using these electrode locations reliable and consistent data, which could be used in clinical applications, cannot be obtained.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...