Dynamic model of amperometric biosensors. Characterisation of glucose biosensor output
Abstract:
An integrated model for the characterisation of the output signal course of oxidase-bound amperometric biosensors is presented and evaluated in the case of glucose biosensors. This model integrates two earlier proposed models, the model of oxygen transducer-based biosensors, allowing the prediction of steady state parameters from the transient response and the dynamic signal lag model, characterising the electrochemical diffusion-limited sensors. The integrated model allows the characterisation of the whole biosensor signal output, originating from the output curve itself with errors less than 3% and no need to determine the system's geometrical, diffusion and partition parameters.