Heliox delivery with noninvasive positive pressure ventilation: a laboratory study.
S Chatmongkolchart1, R M Kacmarek, D R Hess
1Department of Respiratory Care, Massachusetts General Hospital, 55 Fruit Street, Boston, MA 02114, USA.
Respiratory Care
|March 23, 2001
Summary
Helium-oxygen mixture (heliox) delivery during noninvasive positive pressure ventilation (NPPV) depends heavily on heliox flow rate. Some ventilators may malfunction when using heliox, requiring further clinical investigation.
Area of Science:
- Respiratory therapy
- Biomedical engineering
- Gas dynamics
Background:
- Clinical interest exists in using heliox (helium-oxygen mixture) with noninvasive positive pressure ventilation (NPPV).
- Previous reports on delivering heliox with NPPV ventilators are lacking.
- This study evaluates helium concentration ([He]) using heliox with five different NPPV ventilators.
Purpose of the Study:
- To determine the factors affecting helium concentration ([He]) when using heliox with NPPV ventilators.
- To assess the performance and potential malfunctions of NPPV ventilators when delivering heliox.
- To investigate the impact of heliox on positive airway pressure during NPPV.
Main Methods:
- Five NPPV ventilators were tested using a simulated lung model with a standard leak.
- Heliox and oxygen flows were titrated at proximal and distal positions.
- The BiPAP Vision ventilator was also tested using a blender method.
- Ventilators operated in spontaneous/timed mode across various pressure settings.
Main Results:
- Heliox flow, NPPV settings, infusion site, and ventilator type significantly impacted [He].
- Heliox concentrations exceeding 60% were achieved with 18 L/min heliox flow under specific conditions.
- Some ventilators (BiPAP S/T-D30, Quantum, BiPAP Vision) exhibited erratic function.
- Positive airway pressure remained largely unaffected when ventilators functioned correctly.
Conclusions:
- Heliox flow rate is the primary determinant of [He] during NPPV.
- Potential for ventilator malfunction exists when using heliox with NPPV.
- Clinical implications of these findings require further investigation.