Initial experience with a steerable, phased vector array ultrasound catheter in the GI tract
I D Norton1, C J Bruce, J B Seward
1Divisions of Gastroenterology and Cardiology, Mayo Clinic and Foundation, Rochester, MN 55905, USA.
A new ultrasound (US) catheter offers high-resolution imaging and Doppler capabilities for gastrointestinal (GI) tract examinations, potentially increasing endoscopic US (EUS) accessibility. This technology demonstrates feasibility for visualizing GI structures and vessels.
Area of Science:
- Gastroenterology
- Medical Imaging
- Endoscopic Ultrasound
Background:
- Endoscopic ultrasound (EUS) requires substantial capital investment, limiting its widespread use.
- A catheter-based system with phased array scanning and Doppler could enhance EUS accessibility.
- Previous applications of this technology were limited to intracardiac use, not the GI tract.
Purpose of the Study:
- To assess the feasibility of a novel phased array scanning ultrasound (US) catheter for gastrointestinal (GI) tract imaging.
- To evaluate the imaging characteristics and diagnostic capabilities of this new US catheter technology in the GI tract.
Main Methods:
- A 10F, 100 cm torquable catheter with a phased vector array transducer (5.5-10 MHz) was used in anesthetized swine.
- The catheter was inserted into the upper GI tract via a flexible sigmoidoscope, with acoustic coupling achieved using water.
- Needle visualization and Doppler imaging experiments were conducted.
Main Results:
- The catheter facilitated easy acoustic coupling and provided high-resolution imaging of esophageal and stomach walls.
- Excellent tissue resolution and Doppler imaging were achieved up to 10 cm.
- A 22-gauge needle was visualized beyond 4 cm depth, and vascular flow was detected using Doppler.
Conclusions:
- The 10F array US catheter provides high-resolution 2D and Doppler imaging of the GI tract.
- This technology shows promise for new applications in GI diagnostics.
- The system's capabilities may improve the accessibility and utility of endoscopic ultrasound.