Related Experiment Video
Updated: Oct 8, 2026

Effects of Surgical Masks on Cardiopulmonary Function in Healthy Subjects
Published on: February 12, 2021
Validating facemask use for gas exchange analysis in patients with congestive heart failure
D A Baran1, E Rosenwinkel, D K Spierer
1Mt. Sinai Medical Center, Department of Medicine, Division of Cardiology, 1 Gustave L. Levy Place, Box 1030, New York, NY 10029, USA.
Purpose:
Functional limitation, often associated with congestive heart failure (CHF), is most accurately assessed with measurement of peak VO2 (VO2peak) during exercise testing. VO2peak strongly predicts survival in CHF patients and is used to triage patients for cardiac transplantation. However, the traditional mouthpiece/noseclip equipment used for gas exchange analysis is often cumbersome and uncomfortable. Several alternative facemask designs tested previously in healthy cohorts were inferior to the mouthpiece for measuring VO2peak. In this study, a series of facemasks designed specifically for VO2peak testing (Hans-Rudolph) was employed. Although previously validated in healthy young adults, this apparatus has not been validated in patients with severe CHF.
Methods:
The Hans-Rudolph facemasks were compared with the mouthpiece apparatus in 30 patients with severe CHF. Each subject completed a cardiopulmonary exercise test with each device. Twelve subjects were randomized to evaluate a possible training effect. To enhance recruitment, 18 additional subjects were recruited after they had completed a test with the mouthpiece.
Results:
There was no evidence of a training effect and no differences in gas exchange measurements were found between the two devices. The weight-adjusted VO2peak measurements were 17.1 +/- 5.3 ml O2/kg/min for the mouthpiece group and 17.3 +/- 5.6 ml O2/kg/min for the mask group (P = 0.54).
Conclusions:
The facemask was shown to be equivalent to the mouthpiece for measuring VO2peak in CHF patients. Concerns of hypoventilation and decreased VO2peak associated with previous facemask designs were not substantiated. Since successful exercise testing depends on maximal exertion, providing a choice of equipment may facilitate cooperation without sacrificing accuracy.
More Related Videos
Related Concept Videos
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...