3D Ultrasound Imaging: Fast and Cost-effective Morphometry of Musculoskeletal Tissue
Published on: November 27, 2017
Real-time sonography to estimate muscle thickness: comparison with MRI and CT
A C Dupont1, E E Sauerbrei, P V Fenton
1Department of Physiology, Queen's University and Kingston General Hospital, Kingston, Ontario K7L 3N6, Canada.
Purpose:
We investigated the feasibility of using real-time sonography to measure muscle thickness. Clinically, this technique would be used to measure the thickness of human muscles in which intramuscular microstimulators have been implanted to treat or prevent disuse atrophy.
Methods:
Porcine muscles were implanted with microstimulators and imaged with sonography, MRI, and CT to assess image artifacts created by the microstimulators and to design protocols for image alignment between methods. Sonography and MRI were then used to image the deltoid and supraspinatus muscles of 6 healthy human subjects.
Results:
Microstimulators could be imaged with all 3 methods, producing only small imaging artifacts. Muscle-thickness measurements agreed well between methods, particularly when external markers were used to precisely align the imaging planes. The correlation coefficients for sonographic and MRI measurements were 0.96 for the supraspinatus and 0.97 for the deltoid muscle. Repeated sonographic measurements had a low coefficient of variation: 2.3% for the supraspinatus and 3.1% for the deltoid muscle.
Conclusions:
Real-time sonography is a relatively simple and inexpensive method of accurately measuring muscle thickness as long as the operator adheres to a strict imaging protocol and avoids excessive pressure with the transducer.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Magnetic Resonance Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...