Shielding effects of metallic encapsulations and radiographic contrast agents for catheter-based intravascular
1Department of Therapeutic Radiology, Yale University School of Medicine, P.O. Box 208040, New Haven, CT 06520-8040, USA.
Purpose/Objective:
Both photon- and beta-emitting radionuclides for intravascular brachytherapy (IVB) are under active investigation for prevention of restenosis following conventional angioplasty with or without stents. High atomic number materials are usually present in the coronary vessels undergoing treatment in the form of metallic encapsulations, stents, calcified plaque, or radiographic contrast agent. The high atomic number materials are likely to interfere with the photons and betas and, thus, change the dosimetry in the treatment volume. The purpose of this study is to investigate the shielding effects caused by the presence of high atomic number materials in IVB.
Materials And Methods:
Dose rates at various distances in water, with and without the presence of various high atomic number materials, were calculated using Monte Carlo simulation techniques for photon and electron transport in extended media. The high atomic number materials investigated included titanium, stainless steel, calcified plaque, Hypaque, and Omnipaque. A wide range of monoenergetic photon and electron sources and several photon- and beta-emitting radionuclides, which have been under consideration for IVB, were used. The energy of the monoenergetic photon sources was in the range from 10 keV to 1 MeV, and that of the monoenergetic electron sources in the range from 0.5 to 2 MeV. Photon-emitting radionuclides (192)Ir, (125)I, and (103)Pd and beta-emitting radionuclides (90)Y, (32)P, and (188)Re were also considered.
Results:
It was found that the high atomic number materials interfere considerably with the transport of photons of relatively low energies (below 40 keV) and all electron sources. When the energy of photon exceeds 100 keV, the interference becomes minimum for the high atomic number materials that are likely to be present in clinical situations. The shielding correction factors (SCFs) for dose rate at 2 mm from center were essentially 1.00 for photon energies above 100 keV; as the energy decreased below 100 keV, the SCF became smaller reaching a value of almost 0 for the lowest energy studied, 10 keV. For the photon source of (192)Ir, the SCF was essentially 1.00; while for the photon sources of (125)I and (103)Pd, shielding corrections were considerably lower than 1.00 depending on the type and thickness of the high atomic number material. For the beta emitting sources, the shielding effect can be expressed as a loss in effective penetration depth. This loss depends both on the material and its thickness. For titanium and stainless steel, the loss of range was about two and four times the thickness of the metal.
Conclusions:
The effects of high atomic number materials, such as metallic stents, calcified plaque, and contrast agents are minimal for high energy photon emitters, such as (192)Ir. The effects are pronounced for beta emitters and low energy photon emitters, and must be included in dosimetry planning.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography