Related Experiment Video
Updated: Aug 8, 2026

Clean Sampling and Analysis of River and Estuarine Waters for Trace Metal Studies
Published on: July 1, 2016
Determination of picomolar levels of iron in seawater using catalytic cathodic stripping voltammetry
1Oceanography Laboratories, University of Liverpool, UK.
Abstract:
A new procedure for the direct determination of picomolar levels of iron in seawater is presented. Cathodic stripping voltammetry (CSV) is preceded by adsorptive accumulation of the iron(III)-2,3-dihydroxynaphthalene (DHN) complex from seawater, containing 20 microM DHN at pH 8.0, onto a static mercury drop electrode, followed by reduction of the adsorbed species. The reduction current is catalytically enhanced by the presence of 20 mM bromate. Optimized conditions include a 60-s adsorption period at -0.1 V and a voltammetric scan using sampled dc modulation at 10 Hz. In these conditions, a detection limit of 13 pM iron in seawater was achieved which can be lowered further by extending the adsorption time to 300 s. The new catalytic CSV method is approximately 5 times more sensitive than existing CSV methods and was tested on samples from the Atlantic Ocean.
Related Concept Videos
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential ensures...
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...
Voltammetry: Factors Affecting Measurements
Voltammetric Techniques: Linear-Scan (E vs Time)
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Electrodeposition