A comparative study on four screw-plate locking systems in sheep: a clinical and radiological study
C Lindqvist1, A L Söderholm, A Salo
1Department of Oral and Maxillofacial Surgery, Helsinki University Central Hospital, Finland. christian.lindqvist@helsinki.fi
International Journal of Oral and Maxillofacial Surgery
|June 19, 2001
Summary
This study evaluated four mandibular screw-plate systems in sheep, finding that while animals could eat normally, plate and screw fractures occurred in most systems, indicating internal design flaws.
Area of Science:
- Veterinary Surgery
- Biomaterials Science
- Orthopedic Research
Background:
- Mandibular defects pose significant challenges in reconstructive surgery.
- Locking screw-plate systems are commonly used for mandibular reconstruction.
- Evaluating the biomechanical stability and failure modes of these systems is crucial.
Purpose of the Study:
- To assess the clinical and radiological performance of four different mandibular screw-plate locking systems.
- To identify potential failure points, such as plate or screw fractures and bone resorption, in these systems.
- To compare the efficacy and limitations of the tested mandibular reconstruction systems.
Main Methods:
- Surgical creation of 3-4 cm mandibular defects in sheep.
- Reconstruction of defects using four distinct pre-bent angular reconstruction plate systems.
- Fixation with four screws in the mandibular body and three in the ramus for each plate type.
- Clinical and radiological evaluation at 1 and 2 months post-surgery.
Main Results:
- All sheep maintained normal eating and rumination post-operation.
- Fractures occurred in 6 out of 16 plates and 5 out of 112 screws across tested systems.
- One specific plate system exhibited no fractures.
- Slight to moderate bone resorption under the plate was observed in 6 of 16 mandibles.
Conclusions:
- The tested mandibular screw-plate locking systems demonstrate internal shortcomings.
- While functionally adequate for basic functions, mechanical failures (fractures) limit their long-term reliability.
- Further improvements in the design of mandibular reconstruction plates are warranted based on these findings.