Related Experiment Video
Updated: Jul 24, 2026

Intrauterine Telemetry to Measure Mouse Contractile Pressure In Vivo
Published on: April 6, 2015
A new device for measuring intrauterine temperature
A C Sciscione1, T Zainia, T Leet
1Division of Maternal-Fetal Medicine, Department of Obstetrics and Gynecology, Christiana Care Health Services, Newark, Delaware, USA.
Objective:
We evaluated a new device that uses the intrauterine pressure catheter to measure the maternal temperature in patients who are in labor.
Study Design:
The study was conducted at two medical centers, Christiana Hospital in Newark, Delaware, and Saint Louis University/St Mary's Health Center in St Louis, Missouri, from September 1, 1997, to May 2, 1998. An intrauterine pressure catheter with a thermistor sensor in the tip was placed into the uterus after spontaneous rupture of membranes. The intrauterine, oral, and tympanic temperatures were simultaneously obtained immediately after insertion of the intrauterine pressure catheter and then hourly until delivery or the initiation of amnioinfusion.
Results:
The study comprised 97 patients and 404 temperature readings with a temperature range of 34.7 degrees C to 40.7 degrees C. The normal mean +/- SD for the oral, tympanic, and intrauterine temperatures was 36.7 degrees C +/- 0.5 degrees C, 36.8 degrees C +/- 0.5 degrees C, and 37.3 degrees C +/- 0.4 degrees C, respectively. There was a linear relationship among the oral, tympanic, and intrauterine temperatures. All three methods showed a significant increase in mean body temperature after epidural anesthesia.
Conclusion:
The new device, the intrauterine pressure-temperature catheter, provides a convenient and accurate means of continuously measuring uterine temperature in patients who are in labor and require intrauterine monitoring.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...