"Linear" and "derived" otoacoustic emissions in newborns: a comparative study
G Tognola1, P Ravazzani, E Molini
1CNR Centre of Biomedical Engineering, Polytechnic of Milan, Italy.
Ear and Hearing
|June 21, 2001
Summary
The derived averaging technique for click-evoked otoacoustic emissions (CEOAEs) in newborns produced noisier results than the linear technique. This difference in noise levels can impact screening outcomes, affecting the signal-to-noise ratio (SNR) and classification of results.
Area of Science:
- Audiology and Otoacoustic Emissions
- Neonatal Hearing Screening
- Signal Processing in Bioacoustics
Background:
- Click-evoked otoacoustic emissions (CEOAEs) are a non-invasive measure of cochlear outer hair cell function.
- Acquisition techniques, specifically averaging methods, can influence the reliability and interpretation of CEOAE measurements.
- Understanding these influences is crucial for accurate neonatal hearing screening.
Purpose of the Study:
- To compare the effects of two averaging techniques (linear and derived) on CEOAE evaluation in newborns.
- To assess how averaging method impacts signal-to-noise ratio (SNR), reproducibility, and screening pass/fail classification.
Main Methods:
- CEOAEs were recorded from 89 newborns using an Otodynamic ILO88 system.
- Two averaging modes were applied: linear (classic) and derived (artifact cancellation).
- CEOAEs were analyzed for waveform similarity, signal/noise levels, inter-test reproducibility, and pass/fail status based on clinical criteria.
Main Results:
- Derived CEOAEs exhibited higher residual noise (6 dB) and lower inter-test reproducibility compared to linear averaging.
- Waveform differences and time shifts were noted at longer latencies between the two methods.
- Averaging technique significantly influenced pass/fail scoring, with 14% of linear 'pass' results becoming 'fail' with derived averaging.
Conclusions:
- The derived averaging technique results in noisier CEOAEs in newborns compared to the linear technique.
- Screening criteria relying on SNR can be affected by the chosen averaging method.
- Careful consideration of the averaging technique is necessary for consistent and accurate neonatal hearing screening.