Radiation dose monitoring in a breast cancer patient with a pacemaker: a case report
J R Nibhanupudy1, M A de Jesus, M Fujita
1Department of Radiation Oncology, College of Medicine, Howard University Hospital, Washington, DC 20060, USA.
Abstract:
A pacemaker-bearing patient with left-sided breast cancer was treated with adjuvant external beam radiation therapy to the intact breast. She was treated via tangential fields and a single anterior supraclavicular field using 6-MV x-rays. The pacemaker, originally in the treatment field, was removed and a new one placed 4 cm outside the radiation field prior to treatment. Silicon diode chamber Keithley-Farmer type 0.6 cc ionization chamber, and lithium fluoride (LiF) (TLD) chips were used to measure, in vivo, the dose to the pacemaker. From all the fields treated, total dose to the pacemaker was 164 cGy by diode measurements, 182 cGy by ionization chamber measurements, and 171 cGy by TLD measurements. The pacemaker functioned normally throughout the course of treatment.
More Related Videos
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Holter Monitor: 24-Hour Monitoring