Thorium and iodine memory effects in inductively-coupled plasma mass spectrometry
A Al-Ammar1, E Reitznerová, R M Barnes
1Department of Chemistry, University of Massachusetts, Amherst 01003-9336, USA.
Abstract:
Thorium and iodine memory effects have been characterized experimentally for inductively-coupled plasma mass spectrometry by adding ammonia gas directly to the spray chamber and nebulizing aqueous ammonia sample solutions to assess analyte memory retention sites. Thorium memory effect originates from the tendency of an unidentified thorium compound to volatilize from the spray-chamber walls, and not from Th compound adsorption to nebulizer tubing. The mass spectrometer skimmer and sampler cones, ion optics, quadrupole, and other components are not responsible for the memory effect. Unlike that of thorium the iodine memory effect originates from adsorption of iodine compounds on nebulizer tubing surfaces and from volatilization of HI and I2 from the spray-chamber walls. Addition of ammonia sample solutions or ammonia gas directly to the spray chamber eliminated the Th and I memory effects in practical analyses, and blank levels were achieved after 2 min wash-outs. Quantitative recoveries were obtained for Th and I in reference materials.
Related Concept Videos
Mass Spectrometry: Isotope Effect
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Emission Spectroscopy: Interference
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Lab