Electrogram-guided radiofrequency catheter ablation of atrial tissue comparison with thermometry-guide ablation:
D Schwartzman1, J J Michele, C T Trankiem
1Electrophysiology Research Laboratory, Allegheny University of the Health Sciences, Philadelphia, PA, USA. schwartzmand@upmc.edu
Objectives:
To characterize a new method for radiofrequency energy titration during ablation of atrial tissue based on reduction in electrogram amplitude. To compare this method with energy titration using electrode thermometry.
Background:
Complications associated with "anatomy-based" atrial endocardial radiofrequency ablation for suppression of atrial fibrillation may be due to flawed methods of energy titration.
Methods:
The effect of radiofrequency ablation on electrogram amplitude was characterized in a porcine model. A method for energy titration guided by electrogram amplitude reduction ("electrogram-guided") was developed and validated prospectively. Focal (smooth and trabeculated endocardial areas) and linear (smooth endocardial areas) ablation was performed comparing energy titration guided by amplitude reduction with electrode thermometry.
Results:
Amplitude reduction during radiofrequency application was not necessarily equal among unipolar and bipolar electrograms in the ablation region; specific patterns of reduction could be discerned, based on factors such as catheter-endocardial orientation. A criterion of >90 % reduction of unipolar and/or bipolar amplitude best predicted pathologic lesion success. Electrogram-guided focal and linear lesions in smooth areas were free of lesion complications such as endocardial charring, barotrauma, or damage to contiguous extraatrial structures. However, there was a significant incidence of insufficient lesion size, principally non-transmurality, probably due to undertitration of energy. Thermometry-guided focal and linear lesions in smooth areas were uniformly transmural but frequently evidenced complications, due to overtitration of energy. Electrogram-guided focal lesions in trabeculated areas could usually not be achieved, probably due to insufficient contact of the ablation electrode with adjacent pectinate muscles. Thermometry-guided focal lesions in trabeculated areas were smaller than electrogram-guided lesions and did not evidence complications.
Conclusions:
Electrogram-guided lesions in smooth endocardial areas were uncomplicated but had a significant incidence of non-transmurality. Thermometry-guided lesions were uniformly transmural but were frequently complicated.
More Related Videos
06:25Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...