Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor
J M Wright1, G S Mattu, T L Perry
1Department of Pharmacology & Therapeutics, University of British Columbia, Vancouver, BC, Canada. jmwright@interchange.ubc.ca
Background:
To test the accuracy of a new algorithm for the BPM-100, an automated oscillometric blood pressure (BP) monitor, using stored data from an independently conducted validation trial comparing the BPM-100(Beta) with a mercury sphygmomanometer.
Design:
Raw pulse wave and cuff pressure data were stored electronically using embedded software in the BPM-100(Beta), during the validation trial. The 391 sets of measurements were separated objectively into two subsets. A subset of 136 measurements was used to develop a new algorithm to enhance the accuracy of the device when reading higher systolic pressures. The larger subset of 255 measurements (three readings for 85 subjects) was used as test data to validate the accuracy of the new algorithm.
Methods:
Differences between the new algorithm BPM-100 and the reference (mean of two observers) were determined and expressed as the mean difference +/- SD, plus the percentage of measurements within 5, 10, and 15 mmHg.
Results:
The mean difference between the BPM-100 and reference systolic BP was -0.16 +/- 5.13 mmHg, with 73.7% < or = 5 mmHg, 94.9% < or = 10 mmHg and 98.8% < or = 15 mmHg. The mean difference between the BPM-100 and reference diastolic BP was -1.41 +/- 4.67 mmHg, with 78.4% < or = 5 mmHg, 92.5% < or = 10 mmHg, and 99.2% < or = 15 mmHg. These data improve upon that of the BPM-100(Beta) and pass the AAMI standard, and 'A' grade BHS protocol.
Conclusion:
This study illustrates a new method for developing and testing a change in an algorithm for an oscillometric BP monitor utilizing collected and stored electronic data and demonstrates that the new algorithm meets the AAMI standard and BHS protocol.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure