A reevaluation and discussion on the threshold limit for hemolysis in a turbulent shear flow
1Department of Water Resources, and Environmental Engineering, Tamkang University, Taipei 251, Taiwan, R.O.C. lupc@mail.tku.edu.tw
One major factor of red blood cell damage induced by artificial heart valves is the magnitude of turbulent shear stresses in the flow field. An often-cited threshold for hemolysis is 400 N/m(2) (Sallam and Hwang, Biorheology 21 (1984) 783). This value, however, was measured with a one-component laser Doppler anemometer and was not calculated from the major principal Reynolds shear stress, therefore underestimating the threshold. This study performed flow field measurements under similar conditions, using a two-component laser Doppler anemometer and determining major principal Reynolds shear stress, to reevaluate the hemolytic threshold as 800 N/m(2) with an exposure time of 1 ms. The Kolmogorov length scales, approximately 9 microm, are capable of inflicting direct damage to the red blood cells. These results will serve as a more accurate reference in future heart valve design and testing.
One major factor of red blood cell damage induced by artificial heart valves is the magnitude of turbulent shear stresses in the flow field. An often-cited threshold for hemolysis is 400 N/m(2) (Sallam and Hwang, Biorheology 21 (1984) 783). This value, however, was measured with a one-component laser Doppler anemometer and was not calculated from the major principal Reynolds shear stress, therefore underestimating the threshold. This study performed flow field measurements under similar conditions, using a two-component laser Doppler anemometer and determining major principal Reynolds shear stress, to reevaluate the hemolytic threshold as 800 N/m(2) with an exposure time of 1 ms. The Kolmogorov length scales, approximately 9 microm, are capable of inflicting direct damage to the red blood cells. These results will serve as a more accurate reference in future heart valve design and testing.
More Related Videos
09:38A Microfluidic Flow Chamber Model for Platelet Transfusion and Hemostasis Measures Platelet Deposition and Fibrin Formation in Real-time
Published on: February 14, 2017
10:25Microfluidic Flow Chambers Using Reconstituted Blood to Model Hemostasis and Platelet Transfusion In Vitro
Published on: March 19, 2016