A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Right atrial septal electrode for reducing the atrial defibrillation threshold
X Zheng1, M E Benser, G P Walcott
1Division of Cardiovascular Diseases, Department of Medicine, University of Alabama, Birmingham, USA.
Adding a right atrial septum electrode significantly lowers atrial defibrillation threshold energy. This configuration, using the coronary sinus and right atrial septum, requires less energy than the standard right atrial appendage to coronary sinus approach.
Area of Science:
- Cardiovascular Electrophysiology
- Medical Device Technology
Background:
- The standard right atrial appendage (RAA) to coronary sinus (CS) lead configuration for atrial defibrillation has a high energy threshold.
- Previous studies indicated a >50% reduction in atrial defibrillation threshold (ADFT) with a trans-septal electrode.
Purpose of the Study:
- To evaluate if a more clinically practical electrode placement in the right atrium along the atrial septum (RSP) could lower the ADFT.
- Investigate the efficacy of a novel electrode configuration for atrial defibrillation.
Main Methods:
- Sustained atrial fibrillation was induced in sheep using burst pacing and pharmacological agents.
- A custom dual-defibrillation catheter with electrodes in the lateral right atrium (RA), CS, and RSP was used.
- ADFT was assessed for various lead configurations, including RAA-->CS and sequential shocks involving RSP, using biphasic waveforms.
Main Results:
- The sequential-shock configuration CS-->RSP/RA-->RSP significantly lowered delivered energy (0.53+/-0.31 J), peak voltage (86+/-22 V), and peak current (1.6+/-0.6 A) compared to RAA-->CS (1.14+/-0.64 J, 157+/-34 V, 2.5+/-1.1 A).
- No significant difference was found between RAA-->CS and RA-->CS configurations.
- Similar ADFT characteristics were observed for CS-->RSP/RA-->RSP and CS-->RSP/RAA-->RSP configurations.
Conclusions:
- An additional electrode placed at the RSP can markedly reduce the ADFT of the standard RAA-->CS configuration.
- This RSP electrode configuration offers a potentially more effective and lower-energy approach to atrial defibrillation.