Related Experiment Video
Updated: Jul 29, 2026

Electrochemical Preparation of Poly(3,4-Ethylenedioxythiophene) Layers on Gold Microelectrodes for Uric Acid-Sensing Applications
Published on: July 28, 2021
Determination of ammonia in ethylene using ion mobility spectrometry
J H Cross1, T F Limero, J L Lane
1KRUG Life Sciences, Houston, TX 77058, USA. jcross@sdpcmail.jsc.nasa.gov
Abstract:
A simple procedure to analyze ammonia in ethylene by ion mobility spectrometry is described. The spectrometer is operated with a silane polymer membrane., 63Ni ion source, H+ (H2O)n reactant ion, and nitrogen drift and source gas. Ethylene containing parts per billion (ppb) (v/v) concentrations of ammonia is pulled across the membrane and diffuses into the spectrometer. Preconcentration or preseparation is unnecessary, because the ethylene in the spectrometer has no noticeable effect on the analytical results. Ethylene does not polymerize in the radioactive source. Ethylene's flammability is negated by the nitrogen inside the spectrometer. Response to ammonia concentrations between 200 ppb and 1.5 ppm is near linear, and a detection limit of 25 ppb is calculated.
Related Concept Videos
Mass Spectrometry of Amines
Ion-Exchange Chromatography
Atomic Emission Spectroscopy: Instrumentation
Atomic Emission Spectroscopy: Lab