Related Experiment Video
Updated: Oct 4, 2026

Electrochemical Preparation of Poly(3,4-Ethylenedioxythiophene) Layers on Gold Microelectrodes for Uric Acid-Sensing Applications
Published on: July 28, 2021
Direct electrochemical determination of pyruvate in human sweat by capillary zone electrophoresis
1Laboratory of Analytical Science, School of Chemistry, Shandong University, Jinan, PR China.
Abstract:
Capillary zone electrophoresis was employed for the determination of pyruvate in human sweat using electrochemical detection with a carbon fiber microdisk bundle electrode at a constant potential of 1.60 V vs. saturated calomel electrode. The optimum separation conditions are 3.6 x 10(-3) mol/L Na2HPO4-1.4 x 10(-3) mol/L NaH2PO4 (pH 7.2) for the buffer solution, and 18 kV for the separation voltage. The limits of detection of pyruvate are 8.0 x 10(-6) mol/L or 24 fmol (S/N = 3) for the injection voltage of 6 kV and the injection time of 10 s. The relative standard deviation is 2.0% for the migration time and 5.7% for the electrophoretic peak current. The method was applied to determining pyruvate in human sweat.
Related Concept Videos
Electrophoresis: Overview
There...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Capillary Electrophoresis: Instrumentation
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential ensures...
Pyruvate Oxidation
First, the enzyme pyruvate dehydrogenase removes the carboxyl group from pyruvate and releases it as carbon dioxide. The stripped molecule is then oxidized and releases electrons, which are then picked up by NAD+...