Anesthesia breathing circuits protected by the DAR Barrierbac S breathing filter have a low bacterial contamination
D P Vézina1, C A Trépanier, M R Lessard
1Département d'anesthésie, Centre hospitalier universitaire de Québec (Hôpital Enfant-Jésus), Quebec Canada.
Purpose:
In order to reuse the same anesthesia breathing circuit for more than one patient, it has been proposed to add a breathing filter between the Y-piece and the artificial airway. The purpose of this study was to evaluate the in vivo bacterial filtration efficacy of an anesthesia filter in a usual clinical anesthesia setting.
Methods:
A sterile DAR Barrierbac S breathing filter was inserted at the Y-piece of a sterile single-use anesthesia breathing circuit before induction of general anesthesia. At the end of anesthesia, the breathing circuit connector of the filter and of the endotracheal tube connector were cultured separately on growth media (chocolate and blood agar). These were incubated for 48 hr and bacterial identification was conducted using standard methods.
Results:
Bacterial cultures were negative on both sides of the filter membrane of 1842 of the 2001 filters studied. Cultures were positive on the patient side of 104 filters. In two of those, the same bacteria were found on both the circuit side and the patient side of the filter. Therefore these data indicate a clinical effectiveness of 99.9% (confidence interval, CI 95%, 99.6-99.998%), and an in vivo filtration efficacy of 98.08% (CI 95%, 92.54-99.67%).
Conclusion:
Using the upper limit of the CI, it can be assumed that the practice of using a sterile DAR Barrierbac S breathing filter for every patient while reusing the anesthesia breathing circuit would result in a cross contamination rate of the breathing circuit lower than once every 250 cases.
Related Concept Videos
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Cardiopulmonary Resuscitation II: ACLS Airway Management