Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Relationship between carbon number and peak width in gas chromatography
S Lomsugarit1, N Jeyashoke, K Krisnangkura
1Biochemical Technology Division, King Mongkut's University of Technology Thonburi, Toongkru, Bangkok, Thailand. supathral@yahoo.com
Abstract:
When the unadjusted width (wR) is divided into hold-up width (wM) and adjusted width (w'R), a linear relationship between the natural logarithm of the width factor (p') and the carbon number (n) is demonstrated and can be described by: In p' = a + bn where a and b are constants. The coefficients of variance (r2) for n-alkanes, fatty acid methyl esters and fatty alcohols are 0.9997, 0.9996 and 0.9998, respectively.
Related Concept Videos
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Chromatographic Methods: Terminology
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...