Background:
The Cleveland Clinic CorAide left ventricular assist system is based on a small implantable continuous-flow centrifugal blood pump with a completely suspended rotating assembly designed for long-term circulatory support (5 to 10 years).
Methods:
Between June 1999 and August 2000, the CorAide blood pump was implanted in 10 calves for 1 month and in 3 calves for 3 months.
Results:
The mean pump flow and arterial pressure were 6.1 +/- 1.1 L/min and 97 +/- 5 mm Hg, respectively. The mean plasma free-hemoglobin level after postoperative day 3 was 2.0 +/- 1.8 mg/dL. Renal and hepatic function remained normal in all cases. There was no incidence of mechanical failure, hemolysis, bleeding, or systemic organ dysfunction in any of the cases. Significant findings at autopsy were limited to two cases of renal infarction, one of which was associated with an outflow graft infection.
Conclusions:
The CorAide blood pump is easily implanted, reliable, nonhemolytic, and nonthrombogenic, positioning it as a leading third-generation, continuous-flow left ventricular assist system with a completely suspended rotor.
Related Concept Videos
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...