Solid-phase extraction and GC-MS analysis of THC-COOH method optimized for a high-throughput forensic drug-testing
P R Stout1, C K Horn, K L Klette
1Navy Drug Screening Laboratory, Naval Air Station, Jacksonville, Florida 32212, USA.
Journal of Analytical Toxicology
|October 16, 2001
Summary
A new method for analyzing delta9-tetrahydrocannabinol (THC) in urine samples offers rapid, reliable confirmation. This approach significantly reduces lab processing time, waste, and exposure risks for personnel.
Area of Science:
- Forensic Toxicology
- Analytical Chemistry
- Pharmacology
Background:
- Accurate confirmation of delta9-tetrahydrocannabinol (THC) in numerous urine samples is crucial.
- Existing methods may be time-consuming or generate significant waste.
Purpose of the Study:
- To develop an efficient extraction, derivatization, and GC-MS method for THC confirmation in urine.
- To improve laboratory workflow and safety in forensic toxicology.
Main Methods:
- Utilized anion-exchange solid-phase extraction with positive pressure manifold.
- Employed rapid derivatization with pentafluoropropionic anhydride/pentafluoropropanol.
- Analyzed samples using Gas Chromatography-Mass Spectrometry (GC-MS).
Main Results:
- Achieved average recoveries of 95% and a limit of detection of 0.875 ng/mL.
- Demonstrated linearity up to 900 ng/mL THC-COOH using a deuterated internal standard.
- Validated the method on over 1000 human urine samples, confirming excellent agreement with prior results and identifying an improved qualifier ion (m/z 622).
Conclusions:
- The developed method provides a highly efficient and accurate approach for THC confirmation in urine.
- Significant reductions in processing time, waste, and personnel exposure were achieved.
- The method enhances laboratory safety and throughput for forensic drug testing.
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
Gas chromatography–mass spectrometry (GC–MS) is the combination of analytical techniques of gas chromatography and mass spectrometry in a single instrument for analyzing a mixture of compounds. The gas chromatograph separates the compounds in the mixture, and the mass spectrometer analyzes each compound separately to determine the molecular masses and molecular structures.
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Mass Spectrometry: Complex Analysis
Mass spectrometry is an important technique for the identification of pure compounds. However, it has some limitations for the analysis of complex mixtures, often due to excessive fragmentation making the spectrum too complicated to decipher. Mass spectrometry can be combined with suitable separation methods in sequence, forming hyphenated methods, which are useful in the analysis of complex mixtures.
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
High-Performance Liquid Chromatography: Introduction
High-performance liquid chromatography(HPLC), formerly referred to as High-pressure liquid chromatography, is a powerful technique used to separate, identify, and quantify components in complex mixtures. The term "high pressure" refers to using high pressure to push the liquid mobile phase through the tightly packed columns.
In HPLC, two phases play a critical role in the separation process:
In HPLC, two phases play a critical role in the separation process: