Related Experiment Video
Updated: Oct 4, 2026

Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
A breathing circuit disconnection detected by anesthetic agent monitoring
1Department of Anesthesia, Christchurch Hospital, Christchurch, New Zealand.
Purpose:
To describe a case involving a spontaneously breathing patient where a circuit disconnection was detected by a change in monitored anesthetic agent parameters.
Clinical Features:
A patient undergoing shoulder surgery was breathing spontaneously from a circle type anesthesia circuit via a laryngeal mask. A disconnection occurred between the heat and moisture exchanger (HME) and the circle system's Y-piece. As the gas sampling port was integrated into the HME a near normal pattern of CO2 continued to be displayed. The disconnection was noted because of a change in the graphical display of the volatile agent concentration.
Conclusions:
Anesthetic circuit disconnection can be difficult to detect, especially in the spontaneously breathing patient. Capnometry may not detect a disconnection on the machine side of the gas sampling port. Changes in oxygen and volatile agent concentrations may provide an early indication of these types of disconnection.
Related Concept Videos
Inhalational Anesthetics: Overview
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Acute Respiratory Failure-IV
Acute Respiratory Failure-III