Related Experiment Video
Updated: Aug 1, 2026

Thermal Scanning Conductometry (TSC) as a General Method for Studying and Controlling the Phase Behavior of Conductive Physical Gels
Published on: January 23, 2018
Artefacts in multi-echo T2 imaging for high-precision gel dosimetry: III. Effects of temperature drift during
1Radiotherapy and Nuclear Medicine, Ghent University Hospital, Gent, Belgium. Yves.DeDeene@rug.ac.be
Abstract:
In high-precision 3D gel dosimetry, long MR measurement times together with a high amount of RF energy being absorbed by the phantom are very common, and result in a spatially dependent temperature rise in the gel. As T2 of the dosimeter gel is temperature dependent, dose estimation will be affected. In this study we assess the temperature rise in the dosimeter gel by use of MR temperature mapping and computer modelling. It is shown that in conventional MR sequences. where linear k-space sampling is used, a temperature rise of 3 C results in a dose underestimation of 10% over the whole dose map. To correct for these dose errors, a compensation method involving centric k-space ordering is suggested. Computer simulations have been performed to analyse the robustness of the proposed method. Applying the compensated sequence, a temperature rise of 3 C leads to a narrow dose artefact of the order of 3% for a 'worst case' situation in which a single pixel dose gradient is assumed. Negligible deviations are found in the rest of the dose map.
More Related Videos
10:23A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management
Published on: June 23, 2023
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024