Related Experiment Video
Updated: Oct 3, 2026

3D Ultrasound Imaging: Fast and Cost-effective Morphometry of Musculoskeletal Tissue
Published on: November 27, 2017
Rigid registration of 3-D ultrasound with MR images: a new approach combining intensity and gradient information
A Roche1, X Pennec, G Malandain
1INRIA-Projet Epidaure, Sophia Antipolis, France. aroche@sophia.inria.fr
Abstract:
We present a new image-based technique to rigidly register intraoperative three-dimensional ultrasound (US) with preoperative magnetic resonance (MR) images. Automatic registration is achieved by maximization of a similarity measure which generalizes the correlation ratio, and whose novelty is to incorporate multivariate information from the MR data (intensity and gradient). In addition, the similarity measure is built upon a robust intensity-based distance measure, which makes it possible to handle a variety of US artifacts. A cross-validation study has been carried out using a number of phantom and clinical data. This indicates that the method is quite robust and that the worst registration errors are of the order of the MR image resolution.