Transvenous biventricular defibrillation halves energy requirements in patients
C Butter1, E Meisel, J Tebbenjohanns
1Humboldt University and German Heart Institute, Berlin, Germany.christian.butter@charite.de
Background:
Defibrillation thresholds (DFT) with standard implantable cardioverter-defibrillator leads in the right ventricle (RV) may be determined by weak shock field intensity in the myocardium of the left ventricle (LV). Adding a shocking electrode in a coronary vein on the middle of the LV free wall, thereby establishing biventricular defibrillation, substantially reduced defibrillation requirements in animals. We investigated the feasibility of this approach in 24 patients receiving an implantable cardioverter-defibrillator using a prototype over-the-wire temporary LV defibrillation lead.
Methods And Results:
The LV lead was inserted through the coronary sinus, using a guide catheter and guidewire, into a posterior or lateral coronary vein whose location was determined by retrograde venography. Paired DFT testing compared a standard system (RV to superior vena cava plus can emulator [SVC+Can], 60% tilt biphasic shock) to a system including the LV lead. The biventricular system was tested with a dual-shock waveform (20% tilt monophasic shock from LV-->SVC+Can, then 60% tilt biphasic shock from RV-->SVC+Can). Twenty patients completed DFT testing. Venography and LV lead insertion time was 46+/-40 minutes. The biventricular system reduced mean DFT by 45% (8.9+/-1.1 J versus 4.9+/-0.5 J, P<0.001). Twelve patients (60%) had a standard system DFT >/=8 J, and the biventricular system gave a lower DFT in all patients. There were no adverse events related to the use of the LV lead, which was removed after testing.
Conclusions:
Internal defibrillation using a transvenously inserted LV lead is feasible, produces significantly lower DFTs, and seems safe under the conditions tested. Biventricular defibrillation may be a useful option for reducing DFTs or could be added to an LV pacing lead for heart failure.
More Related Videos
09:47A Rat Model of Ventricular Fibrillation and Resuscitation by Conventional Closed-chest Technique
Published on: April 26, 2015
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Half wave rectifier
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation III: AED Use
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias