Sensitivity of spiral computed tomography scanning for detecting intraocular foreign bodies
A B Dass1, P J Ferrone, Y R Chu
1Associated Retinal Consultants, P.C. , R6007 Reichert Health Building, 5333 McAuley Drive, Ypsilanti, MI 48197, USA.
Objective:
The purpose of this study is to determine whether 3-mm computed tomography (CT) cuts are equivalent to 1-mm CT cuts for detecting small intraocular foreign bodies (IOFBs).
Design:
Experimental instrument validation study.
Participants:
Seventy-two porcine eyes were divided into three groups. Each group had 0.5 ml of human blood injected into the vitreous cavity and were surgically implanted with IOFBs.
Intervention:
Each eye was surgically implanted with 0.5-mm metallic, 0.5-mm glass, or 0.5-mm stone IOFBs. Two additional eyes were left without IOFBs to serve as negative controls.
Main Outcome Measures:
Axial CT images of both 3-mm and 1-mm thickness were obtained and analyzed in masked fashion by two separate neuroradiologists.
Results:
The sensitivity for detecting 0.5-mm metallic, 0.5-mm glass, and 0.5-mm stone IOFBs with 3-mm CT images was 100%. The sensitivity for detecting 0.5-mm metallic, 0.5-mm glass, and 0.5-mm stone IOFBs with 1-mm CT images was also 100% (confidence interval, 95%; range, 0.88-1.00). Two negative control eyes without IOFBs were also correctly evaluated by the neuroradiologists.
Conclusions:
With modern spiral CT scanning, 3-mm cuts are as sensitive as 1-mm cuts for detecting small metallic, glass, and stone IOFBs.