Performance evaluation of QuantiBRITE phycoerythrin beads
Cytometry
|December 18, 2001
Summary
QuantiBRITE phycoerythrin (PE) beads offer a reliable method for standardizing antibody bound per cell (ABC) quantitation. This tool ensures consistent results across different laboratories and over time for CD4 and CD69 analysis.
Area of Science:
- Immunology
- Flow Cytometry
- Biotechnology
Background:
- Standardization of antibody quantitation is crucial for reliable flow cytometry analysis.
- Phycoerythrin (PE) beads offer a potential solution for standardizing measurements of antibodies bound per cell (ABC).
Purpose of the Study:
- To evaluate the performance of QuantiBRITE PE beads for standardizing CD4 and CD69 quantitation.
- To assess the precision, inter-instrument variability, and temporal stability of PE bead-based standardization.
Main Methods:
- Whole blood was stained for CD4 quantitation using a no-wash/no-lyse format.
- Whole blood was activated and stained for CD69 quantitation using a lyse-no-wash format.
- QuantiBRITE PE beads were used to standardize antibody bound per cell (ABC) measurements.
Main Results:
- CD4 ABC quantitation showed high precision (<1.0% CV within-assay, <5.0% CV across instruments).
- CD4 ABC quantitation demonstrated good temporal stability (1.3%-3.2% CV within-donor over 2 months).
- CD69 ABC quantitation showed acceptable precision (<3.0% CV within-assay) but higher temporal variability (2%-24% CV within-donor over 1 month).
Conclusions:
- QuantiBRITE PE beads are a valuable tool for standardized flow cytometry analysis across laboratories.
- The beads provide a simple and robust method for quantitating antibody expression levels in terms of ABC when used with 1:1 PE conjugates.