Related Experiment Video
Updated: Aug 6, 2026

09:16
Methods of Soil Resampling to Monitor Changes in the Chemical Concentrations of Forest Soils
Published on: November 25, 2016
Reliability and accuracy of environmental analytical data on moss samples: an interlaboratory comparison
M Dapiaggi1, F Leva, D Rabuffetti
1Joint Research Centre, Environment Institute, Ispra, VA, Italy.
Environmental Technology
|January 5, 2002
Summary
This study compared two laboratories analyzing mineralized moss samples for heavy metals. Results show both labs produced reliable data, with ICP-MS generally outperforming AAS for precision.
Area of Science:
- Environmental Science
- Analytical Chemistry
- Geochemistry
Background:
- Inter-laboratory comparisons are crucial for validating analytical methods.
- Moss is a valuable biomonitor for environmental heavy metal contamination.
- Standardized sample preparation and analysis are essential for accurate results.
Purpose of the Study:
- To assess the reliability of heavy metal (Cd, Cr, Cu, Fe, Ni, Pb) measurements in mineralized moss samples from two independent laboratories.
- To compare the performance of different analytical techniques, including Atomic Absorption Spectrometry (AAS), Inductively Coupled Plasma Mass Spectrometry (ICP-MS), and Instrumental Neutron Activation Analysis (INAA).
- To evaluate the impact of different mineralization procedures on analytical accuracy and precision.
Main Methods:
- Ten moss samples, certified reference materials (CRMs), and blanks were analyzed.
- Independent sample mineralization was conducted in two laboratories.
- Analysis employed AAS, ICP-MS, and INAA for target elements.
- Statistical analyses included linear regression, Student t-test, and maximum likelihood estimation.
Main Results:
- Accuracy tests for AAS showed high precision (RSDs 0.2-2.9%) and values close to CRMs, with minor exceptions.
- Statistical analysis revealed high correlation coefficients (>0.750) and slopes not significantly different from one for most regressions.
- ICP-MS demonstrated superior precision for most elements compared to AAS, with both mineralization methods yielding satisfactory results.
Conclusions:
- Both laboratories demonstrated the capability to produce reliable results for heavy metal analysis in mineralized moss.
- ICP-MS is recommended for higher precision in analyzing most target elements in this matrix.
- The study validates the analytical methodologies and inter-laboratory agreement for environmental monitoring using moss.