[Modification of facial mask on the dead space effect in non-invasive mask ventilation]
Objective:
There were reports concerning the CO2 rebreathing during non-invasive positive pressure ventilation (NIPPV) with full face mask. It is our hypothesis that modification of the mask from one way connection to two ways connection by making a side hole in the mask makes it possible that CO2 inside the mask could be washed out by a constant flow through the mask.
Methods:
A randomized self-control study on CO2 rebreathing was conducted in 7 COPD patients to compare the modified set-up with the conventional one. A BiPAP-30 ventilator and a plateau exhaustion valve (Respironics USA) were employed in the study. In the modified two ways set-up, the exhaustion valve (with distal end blocked) was connected to the side hole of the mask, so that a constant base flow could pass through the mask to the exhaustion valve. The average base flow was 0.43 LPS. The parameters were set as following: S/T mode, f: 15 BPM, pressure support level: 8 cm H2O. Three different levels of PEEP (2, 3 and 5 cm H2O) were used to investigate the PEEP level on CO2 rebreathing. Flow and CO2% were constantly recorded with computer data acquisition and analysis system (Microcal Origin).
Results:
In conventional set-up, there was obvious CO2 rebreathing (rebreathing volume: 83.1 +/- 32.9 ml). In modified connection, the rebreathing volume was only (0.1 +/- 0.4) ml (P < 0.001).
Conclusion:
There was obvious CO2 rebreathing during full face mask NIPPV in conventional set-up. A modified two ways connection could reduce CO2 rebreathing to be near zero, which might be important in the management of hypercapnic respiratory failure with NIPPV.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Cardiopulmonary Resuscitation II: ACLS Airway Management