Related Experiment Video
Updated: Oct 3, 2026

A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
The oxygen concentrator is a suitable alternative to oxygen cylinders in Nepal
Bisharad M Shrestha1, Birendra B Singh, Madhav P Gautam
1Department of Anaesthesia, Bir Hospital, Kathmandu, Nepal. bisharadshrestha@hotmail.com
Purpose:
To review the efficacy and reliability of oxygen concentrators used over the last six years in Nepal. The apparatus used was a DeVilbiss(R) oxygen concentrator that provided O(2) for anesthesia supplemented with compressed air to drive a Penlon Manley Multivent Ventilator(R). It remains difficult to supply oxygen in cylinders to peripheral hospitals in Nepal due to lack of proper roads.
Methods:
We conducted a retrospective analysis of a sample of 378 cases anesthetized at the Bir Hospital and at a private hospital in Kathmandu from April through October 1999. The Bain circuit or its modification was used in adults, and Bain or Ayre's T piece in children. High flows from the oxygen concentrator used with the Bain and Ayre's T-circuits were reduced to 2 L/min, delivered through the halothane vaporizer, supplemented by room air in the modified Bain circuit. Positive pressure ventilation was provided with an Ambubag, Oxford Inflating Bellows or Penlon Manley Multivent Ventilator. Blood pressure, electrocardiogram, FiO(2) and SpO(2) were monitored in all cases.
Results:
Surgery included urologic, general surgery, obstetrics and gynecological procedures, neurosurgery and closed mitral valvotomy. Age ranged from six months to 78 yr. The anesthetic time lasted from 45 min to 12 hr. The FiO(2) ranged from 0.5 to 0.6 in the Bain and Ayre's T circuits, and from 0.34 to 0.40 in the modified Bain circuit with a flow of oxygen of 2 L/min from the concentrator.
Conclusion:
With regular maintenance and servicing done locally, the oxygen concentrator can be used safely in adults and children. Use of the oxygen concentrator is a suitable alternative to oxygen cylinders in the developing world.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue, improving...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...